MODEL XJU-22Z
COMBINED DIAL TYPE CANTILEVER BEAM IZOD IMPACT
TESTER
Introduction
The main technical parameters of
XJU-22Z
Combined cantilever
beam impact tester fully conform to the requirements of ISO 180,
GB/T 1843, GB/T 2611, JB/T 8761 and other standards, and it is
applicable to the determination of impact toughness of non-metallic
materials such as hard plastics, reinforced nylon, glass fiber
reinforced plastics, ceramics, cast stone, electrical insulation
materials, etc. It is a simple structure, convenient operation,
accurate data and reliable testing instrument.
The
impact energy of
this tester is
1J, 2.75J, 5.5J, 11J, and
22J,
the display mode of the instrument is dial pointer type.
Features
-
Combined
impact energy of 1J, 2.75J, 5.5J, 11J and 22J
-
The pointer
has a driving pin and a driven pin. The dial is divided into a
fixed dial and a rotatable digital dial. On the dial, the
outermost layer is the angle, the middle layer is the simple
beam scale, and the inner layer is the cantilever beam scale.
-
The jaw adopts
wedge-shaped self clamping jaw, which can be clamped freely. It
is the latest product newly developed.
Main parameters
Impact speed |
3.5 m/s |
Pendulum energy |
1J, 2.75J, 5.5J,
11J and 22J |
Pendulum torque |
Pd1=0.53590Nm,
Pd2.75=1.47372Nm,
Pd5.5 =2.94744Nm, Pd11=5.8949Nm,
Pd22=11.7898Nm |
The distance between pendulum center and
sample center |
335mm |
Rising angle of the pendulum |
150° |
Blade fillet radius |
R=0.8±0.2mm |
Distance between blade and upper jaw |
22±0.2mm |
Impact blade angle |
75° |
Energy loss of pendulum in no-load impact |
0.5% |
Dial index |
1 / 220 of impact energy |
Net weight |
158 Kg |
Dimensions |
550mm × 350mm ×
850mm |
Package size
|
730mm × 700mm ×
1000mm |
Sample
type:
Sample type |
Length
L (mm) |
Width
b (mm) |
Thickness
h (mm) |
1 |
80±2 |
10.0±0.2 |
4 |
2 |
63.5±2 |
12.7±0.2 |
12.7±0.2 |
3 |
63.5±2 |
12.7±2 |
6.4±2 |
4 |
63.5±2 |
12.7±2 |
3.2±2 |
|
|