|
||||
MODEL DWC-40 LOW TEMPERATURE CHAMBER (FOR IMPACT SAMPLE)
Brief introduction:
Model DWC-40 low temperature chamber is newly researched and developed cooling machine for impact sample according to low temperature device in GB/T229-2007. This device uses overlapping compressor refrigeration technique and thermal equilibrium principle any cycle stir, to make the specimen automatic uniform cooling and constant temperature, which can conform to every temperature-control indicator in the national standard GB/T229-2007 fully.
Main technical specifications:
Cooling range: ﹢30 ~ - 40℃ (room temperature: ≦25) Constant temperature accuracy: <±0.5℃ Effective work space: 240 × 150 × 150mm Specimen quantity: more than 60 pcs., up to 120 pcs. Specimen dimension: 10 × 10 × 55mm Digital timer: 1 sec~99 minutes, resolution 1 sec. Digital display thermometer: resolution 0.1℃ Cooling medium: alcohol or other unfrozen liquid Motor for intermix: 8W Working power supply: 220~240V, 50Hz, 1.0Kw Dimensions: 905 × 510 × 870mm Weight: 60Kg
|
Copyright © All Rights Reserved Laizhou Lyric Testing Equipment Co., Ltd. |