MODEL DWC-196 LOW TEMPERATURE CHAMBER (FOR IMPACT SAMPLE)
Brief introduction:
Model DWC-196 low temperature chamber is the latest refrigeration equipment developed by our company according to the requirements of cryogenic device in GB/T 229-2007 "Metal Charpy Notch Impact Test Method". This equipment adopts liquid nitrogen refrigeration technology, and uses the principle of heat balance to achieve automatic uniform cooling and constant temperature of the sample, which can fully meet the temperature control indicators specified in the national standard GB/T 229-2007. This equipment is easy to operate and has high efficiency. It is the most ideal sample cooling and heat preservation equipment in metal low temperature impact test. It can also be used in other low temperature detection and test work.
Main technical specifications:
Cooling range: room temperature: ﹢30 ~ - 196℃ (room temperature: ≦25) Constant temperature accuracy: ± 2℃ Cooling speed: about 8℃/min Effective work space: 250 × 160 × 120mm Specimen quantity: more than 60 pcs. Specimen dimension: 10 × 10 × 55mm Digital timer: 1 minute ~ 9999 minutes, resolution 1 minute Cooling medium: liquid nitrogen Motor for intermix: 8W Working power supply: 220~240V, 50Hz, 100W Dimensions: Main machine 720 × 600 × 850mm Liquid nitrogen tank 550 × 550 × 1000mm Weight: Main machine 40Kg Liquid tank 35Kg
Main accessories:
Main machine: 1 set Self pressurized stainless steel liquid nitrogen tank: 1 set Solenoid valve: 1 pc. Sensor: 1 pc. Sample baskets: 3 pcs.
|